Skip to content
在CPR反馈设备的帮助下,男人得到了心肺复苏术 在CPR反馈设备的帮助下,男人得到了心肺复苏术

CPRMeter 2

Measure to Improve

在心脏骤停的早期关键时刻,CPRMeter 2通过提供CPR基本参数的实时反馈来帮助急救人员优化CPR性能。

Consistent Quality

Quality compressions - every time

The consensus is clear, high-quality CPR has been shown to save lives.1,2但是,您怎么知道您的团队是否一直在提供高质量的压缩呢?CPRMeter 2是一个简单的工具,您可以用来确保现场所有提供商提供高质量的压缩。3,4The CPRmeter 2 provides real-time measured feedback on depth, rate and release of CPR, while also enabling providers to self-evaluate their performance with event statistics on the spot.

Quick and easy debriefing

Review key performance statistics of delivered compressions with one touch after the event. Resuscitation response scenes tend to be hectic with dispersing responders. Having instant access to objective key performance metrics can lead to quick feedback and debriefing, which has been shown to improve performance and outcomes.6,7,8

By rotating the view within the app, providers can examine visual feedback of their clinical or training performance. This detailed view delivers insights on the essential parameters of CPR. In addition, providers can annotate important events such as obtaining Return of Spontaneous Circulation.

FAQ

CPRMeter 2如何工作?

CPRMeter 2has two embedded sensors: one measuring acceleration and another measuring force. A sophisticated microprocessor continuously measures both of these parameters during each compression, and special algorithms convert the collected data into meaningful information.

The accelerometer measures the depth and rate of chest wall movement during each compression and converts it into distance travelled. The force sensor measures the force applied during CPR and is used for several purposes, the most important being to detect that pressure is not fully released between compressions i.e. provide feedback if the responder is leaning during CPR.

CPRMeter 2utilizes similar Q-CPR technology found in the Philips MRx, Philips XL+ and Philips FR3 defibrillators.

Can CPRmeter 2 be used on infants or children?

CPR feedback devices (CPRmeter) have been shown to improve the quality of CPR, which can ultimately save lives. We are now extending the indication for use of the CPRmeter 2 from >8 years to >1 year in the European market, following the European Guidelines (ERC) for children with Sudden Cardiac Arrest (SCA), so that children can get the same treatment as adults.

因此,有两个CPRMeters 2具有不同的年龄指示来解决不同市场,从而影响用户指南和产品标签。

目录编号和描述

  • 801-00240, CPRmeter 2 (Nordic) > 1 year
  • 801-00241, CPRmeter 2 (Multi) > 1 year
  • 801-00243, CPRmeter 2 (AP) > 8 years
  • 801-00249, CPRmeter 2 (US) > 8 years
  • 801-00245, CPRmeter 2 (Japan) >8 years

在除颤期间,CPRMeter 2可以在患者的胸部施加吗?

When a shock is required the CPRmeter 2 can remain applied to the patient's chest but compressions should be stopped, hands removed from the CPRmeter 2 and remain clear of all patient contact during defibrillation or when otherwise required, in accordance with a proper defibrillation protocol.

What should I do if I am using CPRmeter 2 with a manikin which gives feedback?

You should switch off the feedback given from the manikin. The CPRmeter 2 is a medical device which has been designed to provide the rescuer with CPR guidance for use in a wide range of patients with a range of complex chest properties. CPR manikins tend to have a simplified mechanical model of a chest, which are suitable for training of users in performing CPR. These two models therefore tend to have different measurement technologies and tolerances making it difficult to compare and follow both sets of feedback provided.

我可以在救护车运输期间使用CPRMeter 2吗?

CPRMeter 2is not intended for use in a moving environment, such as an ambulance. If used during patient transport, CPRmeter 2 may provide inaccurate feedback. If CPR is indicated in a moving environment, do not rely on CPRmeter 2 depth feedback during such conditions. It is not necessary to remove the device from the patient.

Can the CPRmeter 2 be used in rain and other adverse weather conditions?

是的,CPRmeter 2有一个IP 55rating.

Protected from limited dust ingress.

Protected from pressure water jets from any direction.

How long will the Battery last?

CPRMeter 2连续监视其2个AAA电池的功率。如果剩余功率少于整个CPR事件所需的电源,则将在心肺复苏术事件中显示低电池指示器。电池应持续至少10集,连续30分钟连续CPR。如果CPRMeter 2保持待机模式,则电池应持续2年,并持续2 30分钟的CPR事件。我们建议使用高质量的碱电池与CPRMeter 2一起使用。

Can I view the Q-CPR Quick Review event statistics after CPRmeter 2 has been switched off?

The CPR event statistics are stored when CPRmeter 2 is turned off. When turned on again, the statistics from the stored CPR event can be reviewed. When CPRmeter 2 is used in a new CPR event, the preceding event’s statistics are removed from the Quick Review and the new event’s statistics are shown in Q-CPR Quick Review.

What is Q-CPR?

Q-CPR是Laerdal开发的商标技术平台,旨在帮助培训提供商在现实生活中提供指南质量CPR。该技术提供了有关CPR参数的客观测量和纠正反馈,以及用于随后汇报或分析的参数记录。

How often should the patient adhesive be replaced?

临床使用后或每两年应更换患者粘合剂。

How can I view the data from CPRmeter 2?

该移动应用程序的创建目的是为用户提供更高细节的汇报,并具有易于访问,存储,共享和可视化的准确数据。

With just a few clicks, you can access in-depth debriefings immediately after events and与您的团队或质量经理分享。

  • You can get an overview of the CPR session in portrait mode
  • Rotate your phone to the side, and geta thorough view on the depth (mm), release, force (kg) and rate of each compression.
  • 捏入和放大以分析特定事件。

Additionally, if you want to share your performance, you will get a .zip folder containing:

  1. A PDF ScoreCard of session performance for debriefing
  2. An excel analytical summary
  3. excel文件中每个压缩的数据

In the early critical moments of a cardiac arrest, the CPRmeter 2 helps first respondersoptimiseCPR performance by providing real-time feedback on essential parameters of CPR.

Features

PHYSICAL

Weight
162G(5.7盎司)
尺寸HXWXD
153 mm x 64 mm x 25 mm
6.0" x 2,5” 1.0”

Ruggedness
IP55 and 1 meter drop test

OPERATION

Display Dimensions
28 mm x 35 mm

Type
TFT显示

解析度
128 x 160 pixels

Display indicators

Low Battery
Small Low battery icon on screen (turning on)
Large Low battery icon on screen (turning off)

Device Error Warning
黄光(固体或闪烁)

所需的服务
Spanner icon

CPR Targets

Compression Depth
≥ 50 mm (2”) ±10 %

压缩释放目标
< 2.5 kg (5.5 lbs)
force: + 1.5kg to -2.0 kg
(+ 3.3 lbs to 4.4 lbs)

压缩率目标
100至120/min±3/min

Compression Counter
1-999 (reset after 5 secs)

Battery

Type
2 x 1.5V AAA

Capacity
至少10集30分钟CPR

Standby life
2年(两年后30分钟CPR)

数据

数据storage
300分钟的数据或20次CPR会话

数据transfer
Bluetooth Smart

ENVIRONMENTAL CONDITIONS

Storage temperature
-20°至70°C(-4°至158°F)

相对湿度
5 % to 75 %

Operating temperature
0° to 50° C (32° to 122° F)

相对湿度
5%至95%

Patient Adhesives

Size
39mm x 90mm(1.5英寸x3.5英寸)

Material
Foam pad with biocompatible adhesive on each side.

保质期
2年应用于CPRMeter或4年未打开的包装

Warranty

One Year limited. Please see Laerdal Global Warranty for terms and conditions

EMC分类
Meets IEC 60601-1-2 and RTCA/DO-160F

参考

1.Steven L. Kronick, Michael C. Kurz, et al Part 4: Systems of Care and Continuous Quality Improvement 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132:S397-S413

2.Nichol G, Thomas E, Callaway CW, Hedges J, Powell JL, Aufderheide TP, Rea T, Lowe R, Brown T, Dreyer J, Davis D, Idris A, Stiell I; Resuscitation Outcomes Consortium Investigators. Regional variation in out-of-hospital cardiac arrest incidence and outcome [published correction appears in JAMA. 2008;300:1763]. JAMA. 2008;300:1423–1431.

3.Halbout Buleon, j . Parienti j j, L。等。2013) AJEM; Improvement in chest compression quality using feedback device (CPRmeter):a simulation randomized crossover study

4.Skorning, M., Beckers, S.K., Brokmann, J.C., et al. (2010), Resuscitation; New Visual Feedback Device Improves Performance of Chest Compressions by Professionals in Simulated Cardiac Arrest”

5.Link MS, Atkins DL, Passman RS, Halperin HR, Samson RA, White RD, Cudnik MT, Berg MD, Kudenchuk PJ, Kerber RE. Part 6: electrical therapies: automated external defibrillators, defibrillation, cardioversion, and pacing: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122(suppl 3):S706 –S719.

6.Edelson,D。P.,B。Litzinger,V。Arora,D。Walsh,S。Kim,D。S. Lauderdale,T。L. Vanden Hoek,L。B. Becker和B. S. Abella。2008年。通过绩效汇报改善医院心脏骤停过程和结果。内科档案168(10):1063-1069。

7.Zebuhr C,Sutton RM,Morrison W,Niles D,Boyle L,Nishisaki A,Meaney P,Leffelman J,Berg RA,Nadkarni VM。小儿心脏骤停后定量汇报的评估。复苏。2012; 83:1124–1128。

8.Dine CJ,Gersh RE,Leary M,Riegel BJ,Bellini LM,Abella BS。通过结合视听反馈和汇报,改善心肺复苏质量和复苏培训。Crit Care Med。2008

Baidu